@ARTICLE{Baszczyk_Mateusz_Silicon_2013, author={Baszczyk, Mateusz and Dorosz, Piotr and Głąb, Sebastian and Kucewicz, Wojciech and Mik, Łukasz and Sapor, Maria}, number={No 4}, journal={Metrology and Measurement Systems}, pages={655-666}, howpublished={online}, year={2013}, publisher={Polish Academy of Sciences Committee on Metrology and Scientific Instrumentation}, abstract={The paper stresses the issue of strong temperature influence on the gain of a Silicon Photomultiplier (SiPM). High sensitivity of the detector to light (single photons) requires stable parameters during measurement, including gain. The paper presents a method of compensating the change of gain caused by temperature variations, by adjusting a suitable voltage bias provided by a precise power module. The methodology of the research takes in account applications with a large number of SiPMs (20 thousand), explains the challenges and presents the results of the gain stabilization algorithm.}, type={Artykuły / Articles}, title={Silicon Photomultiplier Gain Compensation Algorithm in Multidetector Measurements}, URL={http://www.czasopisma.pan.pl/Content/90201/PDF/Journal10178-VolumeXX%20Issue4_12.pdf}, doi={10.2478/mms-2013-0056}, keywords={Silicon photomultiplier, single photons, gain stabilization, temperature influence}, }