@ARTICLE{Neaga_V._Corrosion_2022, author={Neaga, V. and Benea, L.}, volume={vol. 67}, number={No 1}, journal={Archives of Metallurgy and Materials}, pages={371-376}, howpublished={online}, year={2022}, publisher={Institute of Metallurgy and Materials Science of Polish Academy of Sciences}, publisher={Committee of Materials Engineering and Metallurgy of Polish Academy of Sciences}, abstract={Now, the use of any medical device based on metals or alloys, especially intended for dentistry applications, is impossible without preclinical evaluation of its anticorrosion properties. Today, the use of stainless steels with AISI standardization, with predilection 316L and 321, are preferred for ergonomic reasons due to their high operational reliability and optimal mechanical properties for functionality over time. In this regard, 316L and 321 stainless steels are tested for comparison in the solution that simulates human saliva with different pH. Stainless steel samples were subjected to corrosion in Fusayama-Meyer and Carter-Brugirard saliva. In-situ electrochemical measurements were applied, such as the open circuit potential (OCP) and electrochemical impedance spectroscopy (EIS). The results show that the corrosion resistance of 316L is superior to 321 in saliva solution at both pH values.}, type={Article}, title={Corrosion Reactivity in the Pre-Clinical Study of 316L and 321 Stainless Steel for Dentistry Applications}, URL={http://www.czasopisma.pan.pl/Content/122556/PDF-MASTER/AMM-2022-1-49-Benea.pdf}, doi={10.24425/amm.2022.137767}, keywords={Corrosion, electrochemical methods, stainless steel, human saliva}, }