@ARTICLE{Hajiyev_Chingiz_Sensor_2016, author={Hajiyev, Chingiz}, volume={vol. 23}, number={No 3}, journal={Metrology and Measurement Systems}, pages={413-424}, howpublished={online}, year={2016}, publisher={Polish Academy of Sciences Committee on Metrology and Scientific Instrumentation}, abstract={In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM). As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.}, type={Artykuły / Articles}, title={Sensor Calibration Design Based on D-Optimality Criterion}, URL={http://www.czasopisma.pan.pl/Content/106341/PDF/10.15.15mms-2016-0029%20paper%2009.pdf}, doi={10.1515/mms-2016-0029}, keywords={sensor, calibration curve, calibration design, D-optimality criterion, least squares method}, }