Abstract

In many therapeutic applications of a pulsed focused ultrasound with various intensities the finite- amplitude acoustic waves propagate in water before penetrating into tissues and their local heating. Water is used as the matching, cooling and harmonics generating medium. In order to design ultrasonic probes for various therapeutic applications based on the local tissue heating induced in selected organs as well as to plan ultrasonic regimes of treatment a knowledge of pressure variations in pulsed focused nonlinear acoustic beams produced in layered media is necessary. The main objective of this work was to verify experimentally the applicability of the recently developed numerical model based on the Time- Averaged Wave Envelope (TAWE) approach (Wójcik et al., 2006) as an effective research tool for predicting the pulsed focused nonlinear fields produced in two-layer media comprising of water and tested materials (with attenuation arbitrarily dependent on frequency) by clinically relevant axially-symmetric therapeutic sources. First, the model was verified in water as a reference medium with known linear and nonlinear acoustic properties. The measurements in water were carried out at a 25◦C temperature using a 2.25 MHz circular focused (f/3.0) transducer with an effective diameter of 29 mm. The measurement results obtained for 8-cycle tone bursts with three different initial pressure amplitudes varied between 37 kPa and 113 kPa were compared with the numerical predictions obtained for the source boundary condition parameters determined experimentally. The comparison of the experimental results with those simulated numerically has shown that the model based on the TAWE approach predicts well both the spatial-peak and spatial-spectral pressure variations in the pulsed focused nonlinear beams produced by the transducer used in water for all excitation levels complying with the condition corresponding to weak or moderate source-pressure levels. Quantitative analysis of the simulated nonlinear beams from circular transducers with ka ≫ 1 allowed to show that the axial distance at which sudden accretion of the 2nd or higher harmonics amplitude appears is specific for this transducer regardless of the excitation level providing weak to moderate nonlinear fields. For the transducer used, the axial distance at which the 2nd harmonics amplitude suddenly begins to grow was found to be equal to 60 mm. Then, the model was verified experimentally for two-layer parallel media comprising of a 60-mm water layer and a 60-mm layer of 1.3-butanediol (99%, Sigma-Aldrich Chemie GmbH, Steinheim, Germany). This medium was selected because of its tissue-mimicking acoustic properties and known nonlinearity parameter B/A. The measurements of both, the peak- and harmonic-pressure variations in the pulsed nonlinear acoustic beams produced in two-layer media (water/1.3-butanediol) were performed for the same source boundary conditions as in water. The measurement results were compared with those simulated numerically. The good agreement between the measured data and numerical calculations has shown that the model based on the TAWE approach is well suited to predict both the peak and harmonic pressure variations in the pulsed focused nonlinear sound beams produced in layered media by clinically relevant therapeutic sources. Finally, the pulsed focused nonlinear fields from the transducer used in two-layer media: water/castor oil, water/silicone oil (Dow Corning Ltd., Coventry, UK), water/human brain and water/pig liver were predicted for various values of the nonlinearity parameter of tested media.

Go to article

Authors and Affiliations

Tamara Kujawska
Download PDF Download RIS Download Bibtex

Abstract

Image-guided High Intensity Focused Ultrasound (HIFU) technique is dynamically developing technology for treating solid tumors due to its non-invasive nature. Before a HIFU ablation system is ready for use, the exposure parameters of the HIFU beam capable of destroying the treated tissue without damaging the surrounding tissues should be selected to ensure the safety of therapy. The purpose of this work was to select the threshold acoustic power as well as the step and rate of movement of the HIFU beam, generated by a transducer intended to be used in the HIFU ablation system being developed, by using an array of thermocouples and numerical simulations. For experiments a bowl-shaped 64-mm, 1.05 MHz HIFU transducer with a 62.6 mm focal length (f-number 0.98) generated pulsed waves propagating in two-layer media: water/ex vivo pork loin tissue (50 mm/40 mm) was used. To determine a threshold power of the HIFU beam capable of creating the necrotic lesion in a small volume within the tested tissue during less than 3 s each tissue sample was sonicated by multiple parallel HIFU beams of different acoustic power focused at a depth of 12.6 mm below the tissue surface. Location of the maximum heating as well as the relaxation time of the tested tissue were determined from temperature variations recorded during and after sonication by five thermo-couples placed along the acoustic axis of each HIFU beam as well as from numerical simulations. The obtained results enabled to assess the location of each necrotic lesion as well as to determine the step and rate of the HIFU beam movement. The location and extent of the necrotic lesions created was verified using ultrasound images of tissue after sonication and visual inspection after cutting the samples. The threshold acoustic power of the HIFU beam capable of creating the local necrotic lesion in the tested tissue within 3 s without damaging of surrounding tissues was found to be 24 W, and the pause between sonications was found to be more than 40 s.

Go to article

Authors and Affiliations

Łukasz Fura
Tamara Kujawska
Download PDF Download RIS Download Bibtex

Abstract

We have designed and built ultrasound imaging-guided HIFU ablative device for preclinical studies on small animals. Before this device is used to treat animals, ex vivo tissue studies were necessary to determine the location and extent of necrotic lesions created inside tissue samples by HIFU beams depending on their acoustic properties. This will allow to plan the beam movement trajectory and the distance and time intervals between exposures leading to necrosis covering the entire treated volume without damaging the surrounding tissues. This is crucial for therapy safety. The objective of this study was to assess the impact of sonication parameters on the size of necrotic lesions formed by HIFU beams generated by 64-mm bowl-shaped transducer used, operating at 1.08 MHz or 3.21 MHz. Multiple necrotic lesions were created in pork loin samples at 12.6-mm depth below tissue surface during 3-s exposure to HIFU beams with fixed duty-cycle and varied pulse-duration or fixed pulse-duration and varied duty-cycle, propagated in two-layer media: water-tissue. After exposures, the necrotic lesions were visualized using magnetic resonance imaging and optical imaging (photos) after sectioning the samples. Quantitative analysis of the obtained results allowed to select the optimal sonication and beam movement parameters to support planning of effective therapy.
Go to article

Bibliography

1. Chauhan S. (2008), FUSBOTs: image-guided robotic systems for Focused Ultrasound Surgery, Medical Robotics, Vanja Bozovic, I-Tech Education and Publishing, Vienna, Austria.
2. Choi J.W. et al. (2014), Portable high-intensity focused ultrasound system with 3D electronic steering, real-time cavitation monitoring, and 3D image reconstruction algorithms: a preclinical study in pigs, Ultrasonography, 33(3): 191–199, doi: 10.14366/usg.14008.
3. Duck F.A. (1990), Physical Properties of Tissue: A Comprehensive Reference Book, Academic Press, London.
4. Ebbini E.S., ter Haar G. (2015), Ultrasound-guided therapeutic focused ultrasound: current status and future directions, International Journal of Hyperthermia, 31(2): 77–89, doi: 10.3109/02656736.2014.995238.
5. Ellens N. et al. (2015), The targeting accuracy of a preclinical MRI-guided focused ultrasound system, Medical Physics, 42(1): 430–439, doi: 10.1118/1.4903950.
6. Fukuda H. et al. (2011), Hyper-echo in ultrasound images during high-intensity focused ultrasound ablation for hepatocellular carcinomas, European Journal of Radiology, 80(3): e571–e575, doi: 10.1016/ j.ejrad.2011.09.001.
7. Fura Ł., Kujawska T. (2019), Selection of exposure parameters for a HIFU ablation system using an array of thermocouples and numerical simulations, Archives of Acoustics, 44(2): 349–355, doi: 10.24425/ aoa.2019.128498.
8. Guillaumier S. et al. (2018), A multicentre study of 5- year outcomes following focal therapy in treating clinically significant nonmetastatic prostate cancer, European Urology, 74(4): 422–429, doi: 10.1016/j.eururo. 2018.06.006.
9. ter Haar G. (2007), Therapeutic applications of ultrasound, Progress in Biophysics & Molecular Biology, 93(1–3): 111–129, doi: 10.1016/j.pbiomolbio. 2006.07.005.
10. Hand J.W., Shaw A., Sadhoo N., Rajaqopal S., Dickinson R.J., Gavrilov L.R. (2009), A random phased array device for delivery of high intensity focused ultrasound, Physics in Medicine & Biology, 54(19): 5675–5693, doi: 10.1088/0031-9155/54/19/002.
11. Koch T., Lakshmanan S., Brand S., Wicke M., Raum K., Moerlein D. (2011), Ultrasound velocity and attenuation of porcine soft tissues with respect to structure and composition: I. Muscle, Meat Science, 88(1): 51–58, doi: 10.1016/j.meatsci.2010.12.002.
12. Kujawska T., Secomski W., Byra M., Postema M., Nowicki A. (2017), Annular phased array transducer for preclinical testing of anti-cancer drug efficacy on small animals, Ultrasonics, 76: 92–98, doi: 10.1016/j.ultras.2016.12.008.
13. Law W.K., Frizzell L.A., Dunn F. (1985), Determination of the nonlinearity parameter B/A of biological media, Ultrasound in Medicine & Biology, 11(2): 307–318, doi: 10.1016/0301-5629(85)90130-9.
14. Leslie T. et al. (2012), High-intensity focused ultrasound treatment of liver tumours: post-treatment MRI correlates well with intra-operative estimates of treatment volume, The British Journal of Radiology, 85(1018): 1363–1370, doi: 10.1259/bjr/56737365.
15. Li K., Bai J.F., Chen Y.Z., Ji X. (2018), Experimental evaluation of targeting accuracy of an ultrasound- guided phased-array high-intensity focused ultrasound system, Applied Acoustics, 141: 19–25, doi: 10.1016/j.apacoust.2018.06.011.
16. Li S., Wu P.H. (2013), Magnetic resonance imageguided versus ultrasound guided high-intensity focused ultrasound in the treatment of breast cancer, Chinese Journal of Cancer, 32(8): 441–452, doi: 10.5732/cjc.012.10104.
17. Masamune K., Kurima I., Kuwana K., Yamashita H., Chiba T., Dohi T. (2013), HIFU positioning robot for less-invasive fetal treatment, Procedia CIRP, 5: 286-289, doi: 10.1016/j.procir.2013.01.056.
18. Melodelima D., N’Djin W.A., Parmentier H., Chesnais S., Rivoire M., Chapelon J.Y. (2009), Thermal ablation by high-intensity-focused ultrasound using a toroid transducer increases the coagulated volume. Results of animal experiments, Ultrasound in Medicine & Biology, 35(3): 425–435, doi: 10.1016/j.ultrasmedbio.2008.09.020
19. Nassiri D.K., Nicholas D., Hill C.R. (1979), Attenuation of ultrasound in skeletal muscle, Ultrasonics, 17(5): 230–232, doi: 10.1016/0041-624x(79)90054-4.
20. Orsi F., Arnone P., Chen W., Zhang L. (2010), High intensity focused ultrasound ablation: a new therapeutic option for solid tumors, Journal of Cancer Research and Therapeutics, 6(4): 414–420, doi: 10.4103/0973-1482.77064.
21. Schneider C.A., Rasband W.S., Eliceiri K.W. (2012), NIH Image to ImageJ: 25 years of image analysis, Nature Methods, 9(7): 671–675, doi: 10.1038/ nmeth.2089.
22. Shui L. et al. (2015), High-intensity focused ultrasound (HIFU) for adenomyosis: two-year follow-up results, Ultrasonics Sonochemistry, 27: 677–681, doi: 10.1016/j.ultsonch.2015.05.024.
23. Treeby B.E., Jaros J., Rendell A.P., Cox B.T. (2012), Modeling nonlinear ultrasound propagation in heterogeneous media with power law absorption using a k-space pseudo-spectral method, The Journal of the Acoustical Society of America, 131(6): 4324–4336, doi: 10.1121/1.4712021.
24. Veereman G. et al. (2015), Systematic review of the efficacy and safety of high-intensity focused ultrasound for localized prostate cancer, European Urology Focus, 1(2): 158–170, doi: 10.1016/j.euf.2015.04.006.
25. Wang Y., Wang Z.B., Xu Y.H. (2018), Efficacy, efficiency, and safety of magnetic resonance-guided highintensity focused ultrasound for ablation of uterine fibroids: comparison with ultrasound-guided method, Korean Journal of Radiology, 19(4): 724–732, doi: 10.3348/kjr.2018.19.4.724.
26. Wójcik J., Nowicki A., Lewin P.A., Bloomfield P.E., Kujawska T., Filipczynski L. (2006), Wave envelopes method for description of nonlinear acoustic wave propagation, Ultrasonics, 44: 310–329, doi: 10.1016/j.ultras.2006.04.001.
27. Yu T., Xu C. (2008), Hyperecho as the indicator of tissue necrosis during microbubble-assisted high intensity focused ultrasound sensitivity, specificity and predictive value, Ultrasound in Medicine & Biology, 34(8): 1343–1347, doi: 10.1016/j.ultrasmedbio.2008.01.012.
28. Zavaglia C., Mancuso A., Foschi A., Rampoldi A. (2013), High-intensity focused ultrasound (HIFU) for the treatment of hepatocellular carcinoma: is it time to abandon standard ablative percutaneous treatments?, Hepatobiliary Surgery and Nutrition, 2(4): 184–187, doi: 10.3978/j.issn.2304-3881.2013.05.02.
29. Zhang L., Rao F., Setzen R. (2017), High intensity focused ultrasound for the treatment of adenomyosis: selection criteria, efficacy, safety and fertility, Acta Obstetricia et Gynecologica Scandinavica, 96(6): 707–714, doi: 10.1111/aogs.13159.
30. Zhang X., Li K., Xie B., He M., He J., Zhang L. (2014), Effective ablation therapy of adenomyosis with ultrasound-guided high-intensity focused ultrasound, International Journal of Gynecology & Obstetrics, 124(3): 207–211, doi: 10.1016/j.ijgo.2013.08.022.

Go to article

Authors and Affiliations

Łukasz Fura
1
Wojciech Dera
2
Cezary Dziekoński
2
Maciej Świątkiewicz
3
Tamara Kujawska
1

  1. Department of Ultrasound Institute of Fundamental Technological Research, Polish Academy of Sciences
  2. Department of Theory of Continuous Media and Nanostructures Institute of Fundamental Technological Research, Polish Academy of Sciences
  3. Department of Experimental Pharmacology Mossakowski Medical Research Centre, Polish Academy of Sciences
Download PDF Download RIS Download Bibtex

Abstract

Many therapeutic applications of pulsed focused ultrasound are based on heating of detected lesions which may be localized in tissues at different depths under the skin. In order to concentrate the acoustic energy inside tissues at desired depths a new approach using a planar multi-element annular array transducer with an electronically adjusted time-delay of excitation of its elements, was proposed. The 7-elements annular array transducer with 2.4 MHz center operating frequency and 20 mm outer diameter was produced. All its elements (central disc and 6 rings) had the same radiating area. The main purpose of this study was to investigate thermal fields induced in bovine liver in vitro by pulsed focused ultrasonic beams with various acoustic properties and electronically steered focal plane generated from the annular array transducer used. The measurements were performed for the radiating beams with the 20 mm focal depth. In order to maximize nonlinear effects introducing the important local temperature rise, the measurements have been performed in two-layer media comprising of a water layer, whose thickness was specific for the transducer used and equal to 13 mm, and the second layer of a bovine liver with a thickness of 20 mm. The thickness of the water layer was determined numerically as the axial distance where the amplitude of the second harmonics started to increase rapidly. The measurements of the temperature rise versus time were performed using a thermocouple placed inside the liver at the focus of the beam. The temperature rise induced in the bovine liver in vitro by beams with the average acoustic power of 1W, 2 W and 3 W and duty cycle of 1/5, 1/15 and 1/30, respectively, have been measured. For each beam used the exposure time needed for the local tissue heating to the temperature of 43°C (used in therapies based on ultrasonic enhancement of drug delivery or in therapies involving stimulation of immune system by enhancement of the heat shock proteins expression) and to the temperature of 56°C (used in HIFU therapies) was determined. Two sets of measurements were done for each beam considered. First, the thermocouple measurement of the temperature rise was done and next, the real-time monitoring of dynamics of growth of the necrosis area by using ultrasonic imaging technique, while the sample was exposed to the same acoustic beam. It was found that the necrosis area becomes visible in the ultrasonic image only for beams with the average acoustic power of 3 W, although after cutting the sample the thermo ablated area was visible with the naked eye even for the beams with lower acoustic power. The quantitative analysis of the obtained results allowed to determine the exposure time needed to get the necrosis area visible in the ultrasonic image.

Go to article

Authors and Affiliations

Tamara Kujawska
Andrzej Nowicki
Wojciech Secomski
Kazimierz Krawczyk
Download PDF Download RIS Download Bibtex

Abstract

Therapeutic and surgical applications of focused ultrasound require monitoring of local temperature rises induced inside tissues. From an economic and practical point of view ultrasonic imaging techniques seem to be the most suitable for the temperature control. This paper presents an implementation of the ultrasonic echoes displacement estimation technique for monitoring of local temperature rise in tissue during its heating by focused ultrasound The results of the estimation were compared to the temperature measured with thermocouple. The obtained results enable to evaluate the temperature fields induced in tissues by pulsed focused ultrasonic beams using non-invasive imaging ultrasound technique

Go to article

Authors and Affiliations

Piotr Karwat
Jerzy Litniewski
Tamara Kujawska
Wojciech Secomski
Kazimierz Krawczyk